How to maintain an ozone therapy device?

2025-01-07 110

The service life of ozone therapy device>; Reasonable daily maintenance can extend its service life for 8 years. At the same time, there are certain requirements for the working environment of ozone therapy devices. Unreasonable working environments not only pose safety hazards, but also reduce the service life of ozone therapy devices.

1
Daily maintenance
The daily maintenance of ozone therapy equipment is mainly the responsibility of clinical users, who should keep the surface of the ozone therapy equipment clean, check whether the power supply device is normal before use, observe the function and performance of the equipment during use, and fill in the usage record in a timely manner. When the equipment malfunctions, in addition to making necessary records, professional equipment maintenance personnel should be notified in a timely manner

2
Regular maintenance
In order to ensure the normal use of ozone therapy equipment, regular surface dust removal and cleaning should be carried out according to the performance requirements of the equipment, and performance testing should be conducted to timely check for damaged parts, and to check whether the equipment's voltage stabilization and grounding conditions are good. The detailed preventive maintenance content should include:

Appearance inspection: Firstly, check whether the buttons, switches, connectors, and sockets of the equipment are loose or misplaced, whether the contacts of the plugs and sockets are oxidized, rusted, or poorly connected, whether the power cord is aging, and whether the exhaust is normal.

Cleaning and maintenance: Clean the surface of the equipment, prohibit cleaning the internal electrical and mechanical parts, and disinfect the plugs and sockets related to the equipment to prevent poor contact.

Replacement of vulnerable parts: For components that have reached the end of their service life and performance decline, or components that do not meet the requirements or accessories (such as filters) that require regular replacement as specified in the user manual, they should be replaced in a timely manner to eliminate obvious and potential faults in the equipment, prevent possible faults from expanding or causing overall machine failures.

Function check: Check whether the indicator lights and indicators are normal when turning on the device. By adjusting and setting the various switches and buttons, enter the various function settings to check whether the basic functions of the device are normal.

Performance testing calibration: Conduct necessary calibration and adjustment according to the instructions to ensure that all technical indicators of the equipment meet the standards and ensure the normal operation of the equipment during the treatment process.



Ozone therapy device
Specific requirements for work environment


1. Ozone therapy devices are strictly prohibited from being placed in places with direct sunlight or high relative humidity.
2. Ozone therapy devices are strictly prohibited from being placed near heat sources or in environments with excessively low temperatures. It is recommended to store them at a temperature of -10~60 ℃.
3. Ensure that the ventilation openings of the equipment are unobstructed and not covered.
4. Do not stack or wrap circuit wires randomly.
5. The ozone therapy device is a movable fixed device, please install and connect it according to regulations.
6. Do not use this equipment in rooms where flammable and explosive materials are piled up.
7. When using the ozone therapy device, do not approach other high-frequency devices (such as high-frequency electron emission devices, radiation therapy devices, etc. Although the German Caterpillar ozone therapy device has anti-interference ability, it cannot be ruled out that there is a small probability of signal interference).
8. It is recommended to use the ozone therapy device at a temperature of 10-40 ℃.